Applications:
The Anesthesia Machine is a movable unit integrated with anesthesia and ventilation, which could be applied to closed anesthesia,semi-closed anesthesia, and semi-open anesthesia, and be applicable to Adult / Pediatric inhalationSpecifications:
Display | High light LED display |
Ventilation Volume | 50 ~ 1600ml |
Breath frequency | 1 ~ 90BPM |
I:E ratio | 4:1 ~ 1:6 |
Monitor parameter | Tidal volume, Breath frequency, MV,Airway pressure |
Alarm | MV high alarm, MV low alarm, TV high alarm, TV low alarm |
O2 Flush | 25~75L/min. |
Circle absorber | Compact circle absorber, Safety valve pressure limit: 6 Kpa Good air tightness, high reliability, and minor respiration impedance With Manual/ Ventilation switch |
Gas supply | O2 or O2+ N2O or O2+ Air |
Gas Supply pressure: | O2: 0.25~0.65Mpa N2O(Air): 0.25~0.65Mpa |
Physical specification | Whole machine: Length: 790mm Width: 505 mm Height: 1350mm Weight: 70Kg Wheels: 100mm (diameter), 125mm (height), with brakes on the front two wheels |
Environments | Environment Temperature: -9~+35°C Relative humidity limits: 20%~60% Atmosphere pressure: 700~1060hPa |